6  RLC Swept SA Measurement Front-Panel & SCPI Reference

View/Display

Measurement

Swept SA

Remote Command

:TRACe:DISPlay:VIEW:SPECtrogram:POSition

:TRACe:DISPlay:VIEW:SPECtrogram:POSition?

Example

:TRAC:DISP:VIEW:SPEC:POS 146

Preset

Unaffected by Mode Preset, but set to 0 by Restore Mode Defaults.  The value is remembered
when you go in and out of Spectrogram View.

State Saved

Saved in Instrument State

Min

0

Max

300

Default Unit

Enter

Initial S/W Revision

A.07.01

Full Screen

When Full Screen is pressed the measurement window expands horizontally over the entire
instrument display. The screen graticule area expands to fill the available display area.

It turns off the display of the softkey labels, however the menus and active functions still work.
(Though it would obviously be very hard to navigate without the key labels displayed.) Pressing
Full Screen

again while Full Screen is in effect cancels Full Screen.

Note that the banner and status lines are unaffected. You can get even more screen area for your
data display by turning off the Meas Bar (in the Display menu) which also turns off the settings
panel.

Full Screen is a Meas Global function. Therefore it is cancelled by the Preset key.

Key Path

Display

Remote Command

:DISPlay:FSCReen[:STATe] OFF|ON|0|1

:DISPlay:FSCReen[:STATe]?

Preset

Unaffected by Preset but set to Off  by Restore Misc Defaults or shutdown and restart

State Saved

Not saved in instrument state.

Backwards
Compatibility SCPI

:DISPlay:MENU[:STATe] OFF|ON|0|1

This emulates ESA full screen functionality, which is the same as the FSCReen command in
PSA except that the sense of on/off is reversed (that is, OFF means the menus are OFF, so
Fullscreen is ON) and the default is ON (meaning Fullscreen is OFF).

Remote Language Compatibility Measurement Application Reference