6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Mode
String length is COUNt * 7 – 1. (7 = Model Number length + 1 for comma. –1 = no comma for the
1st entry.)
Preset
Not affected by a Preset
State Saved
Not saved in instrument state.
Initial S/W Revision
Prior to A.02.00
Application Catalog Revision
Returns the Revision of the provided Model Number.
Remote Command
:SYSTem:APPLication:CATalog:REVision?
Example
:SYST:APPL:CAT:REV? ‘N9060A’
Notes
Returned value is a quoted string of revision for the provided Model Number. The revision will be
a null-string ("") if the provided Model Number is not installed and licensed. Example, if SAMS is
installed and licensed:
"1.0.0.0"
Preset
Not affected by a Preset.
State Saved
Not saved in instrument state.
Initial S/W Revision
Prior to A.02.00
Application Catalog Options
Returns a list of Options for the provided Model Number
Remote Command
:SYSTem:APPLication:CATalog:OPTion?
Example
:SYST:APPL:CAT:OPT? ‘N9060A’
Notes
Returned value is a quoted string of a comma separated list of Options, in the same format as
*OPT? or :SYSTem:OPTion?. If the provided Model Number is not installed and licensed a null-
string ("") will be returned. Example, if SAMS is installed and licensed:
"2FP"
String length is a maximum of 255 characters.
Preset
Not affected by a Preset
State Saved
Not saved in instrument state.
Initial S/W Revision
Prior to A.02.00