6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
Trig Slope
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.
Key Path
Trigger, External 2
Remote Command
:TRIGger[:SEQuence]:EXTernal2:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:EXTernal2:SLOPe?
Example
TRIG:EXT2:SLOP NEG
Couplings
This same slope is used in the Ext2 selection for the trigger source in the Trigger menu and for
the period timer sync source (in the Trigger menu and in the Gate Source menu).
Preset
POSitive
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:EXTernal2:SLOPe
Backwards Compatibility
Notes
The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.
Initial S/W Revision
Prior to A.02.00
RF Burst
Pressing this key, when it is not selected, selects the RF Burst as the trigger. A new
sweep/measurement will start when an RF burst envelope signal is identified from the signal at
the RF Input connector. Pressing this key, when it is already selected, accesses the RF Burst
trigger setup menu.
In some models, a variety of burst trigger circuitry is available, resulting in various available
burst trigger bandwidths. The analyzer automatically chooses the appropriate trigger path based
on the hardware configuration and other settings of the analyzer.
Key Path
Trigger
Example
TRIG:SOUR RFB Swept SA measurement
TRIG:
State Saved
Saved in instrument state
Status Bits/OPC
dependencies
The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up