3 Programming the Analyzer
STATus Subsystem
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Integrity Uncalibrated Enable
This command determines which bits in the Questionable Integrity Uncalibrated Condition
Register will set bits in the Questionable Integrity Uncalibrated Event register, which also sets the
Data Uncalibrated Summary bit (bit 3) in the Questionable Integrity Register. The variable
Mode
All
Remote
Command
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle?
Example
STAT:QUES:INT:UNC:ENAB 1
Oversweep (Meas Uncal) will be reported to the Integrity Summary of the Status Questionable register.
Preset
32767
Min
0
Max
32767
Status
Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
Questionable Integrity Uncalibrated Event Query
This query returns the decimal value of the sum of the bits in the Questionable Integrity
Uncalibrated Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:UNCalibrated[:EVENt]?
Example
STAT:QUES:INT:UNC?
166