6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
Coupling
Some probe types allow coupling to reject low frequencies. This will filter out the DC component of
a signal that is composed of a DC bias plus some AC signal. This control is visible only for probe
types that have this capability.
Key Path
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe
Remote Command
:INPut:COUPling:I|Q DC|LFR1|LFR2
:INPut:COUPling:I|Q?
Example
Set the probe to low frequency rejection below 1.7 Hz.
INP:COUP:I LFR1
Notes
Only some probe types support Coupling. For those that do, each probe type has its own Coupling
setting. As probes are changed the Coupling value will reflect the new probe's setting. Changing
the Coupling affects only the current probe type's setting and leaves all others unchanged.
Preset
DC
State Saved
Saved with probe calibration data. It survives a power cycle and is not affected by a Preset or
Restore.
Range
DC | AC 1.7 Hz LFR1 | AC 0.14 Hz LFR2
Readback Text
DC | LFR1 | LFR2
Initial S/W Revision
Prior to A.02.00
DC
Turns off low frequency rejection, allowing signals down to DC.
Key Path
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling
Example
Turn off low frequency rejection on the I channel
INP:COUP:I DC
Initial S/W Revision
Prior to A.02.00
LFR1
Turns on low frequency rejection, rejecting signal component lower than 1.7 Hz.
Key Path
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling
Example
Turn on low frequency rejection on the I channel for frequencies lower than 1.7 Hz
INP:COUP:I LFR1
Initial S/W Revision
Prior to A.02.00