6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trace/Detector

Preset

Trace number – 1 (wraps at 1). For example, for Trace 1, the Second Trace presets to Trace 6; for
Trace 6, it presets to Trace 5.

State Saved

The Second Trace operand for each trace is stored in instrument state

Readback

Trace

Initial S/W Revision

Prior to A.02.00

Operand 2

Selects the second trace operand to be used for the trace math functions for the destination trace.

Key Path

Trace/Detector, Math, Trace Operands

Dependencies

The Second Trace cannot be the same as the destination trace. The destination trace number is
gray on the key, and the underline skips that number when selecting the trace.

Preset

Trace number – 1 (wraps at 1). For example, for Trace 1, the Second Trace presets to Trace 6; for
Trace 6, it presets to Trace 5.

State Saved

The Second Trace operand for each trace is stored in instrument state

Readback

Trace

Initial S/W Revision

Prior to A.02.00

Copy/Exchange

This menu lets you copy any trace to any other trace, or exchange any trace with any other trace.
The action is performed once, it is not an “every sweep” type of thing.

The X-Axis settings and domain of a trace go with it when it is copied or exchanged.

Key Path

Trace/Detector

Remote Command

:TRACe:COPY TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6, TRACE1
| TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6

:TRACe:COPY?

Example

TRAC:COPY TRACE1, TRACE3 Copies Trace 1 to Trace 3 and puts Trace 3 in Update=Off,
Display=On

Notes

The TRACe:COPY command is of the form:
:TRACe:COPY ,

Notes

In the case of a Copy, the destination trace is put in Update=Off, Display=On after the copy. In
the case of an Exchange, both traces are put into Update=Off, Display=On after the exchange.

Remote Language Compatibility Measurement Application Reference

1375