6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
Scope
Meas Global
Remote Command
:TRIGger[:SEQuence]:RFBurst:LEVel:ABSolute
:TRIGger[:SEQuence]:RFBurst:LEVel:ABSolute?
Example
TRIG:RFB:LEV:ABS 10 dBm
sets the trigger level of the RF burst envelope signal to the absolute level of 10 dBm
Notes
Sending this command does not switch the setting from relative to absolute; to switch it you
need to send the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE command, below.
Amplitude Corrections are not taken into account by the Absolute Trigger Level. For example, if
you have given yourself effective gain with an amplitude correction factor, the Absolute Trigger
will not fire until you have set the trigger level that far below the displayed signal level, rather
than simply to the displayed signal level. This is only true for Amplitude Corrections, not
External Gain or Ref Level Offset functions.
If mode is Bluetooth, the default value is –50 dBm.
Couplings
This same level is used for the RF Burst trigger source in the Trigger menu, for the RF Burst
selection in the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu),
and also for the RF Burst selection in the Gate Source menu
Preset
–20 dBm
State Saved
Saved in instrument state
Min
–200 dBm
Max
100 dBm
Default Unit
depends on the current selected Y-Axis unit
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:RFBurst:LEVel:ABSolute
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.04.00
Key Path
Trigger, RF Burst
Remote Command
:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE ABSolute|RELative
:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE?
Example
TRIG:RFB:LEV:TYPE REL
sets the trigger level type of the RF burst trigger to Relative.
Preset
ABSolute
State Saved
Saved in instrument state
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference