I/q guided calibration – Agilent Technologies Signal Analyzer N9030a User Manual
Page 666
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
Key Path
Input/Output, Output Config, Aux IF Out
Example
OUTP:AUX LOGVideo
causes the aux output type to be Fast Log Video
Dependencies
Does not appear unless Option ALV is installed.
The output is off during an alignment but not during a marker count, and is not blanked during
retrace (after a sweep and before the next sweep starts).
Readback Text
Fast Log Video
Initial S/W Revision
A.04.00
I/Q Guided Calibration
Calibrating the Baseband I/Q ports requires several steps and manual connections. The Guided
Calibration interactively guides you through the required steps, displaying diagrams to help with
the connections. The steps vary depending on the setup.
In the Guided Calibration windows, the date and time of the last calibration are displayed. If any
of the items listed are displayed in yellow, this indicates that the calibration for that item is
inconsistent with the latest calibration, and you should complete the entire calibration process
before you exit the calibration.
I/Q Isolation Calibration
The I/Q Isolation Calibration must be run before calibrating any port with either the I/Q Cable
Calibration or I/Q Probe Calibration. This calibration is performed with nothing connected to any
of the front panel I/Q ports. This is the first step in both the I/Q Cable Calibration and the I/Q
Probe Calibration.
I/Q Isolation Calibration Time (Remote Command Only)
Returns the last date and time that the I/Q Isolation Calibration was performed. This is a remote
query command only.
Remote Command
:CALibration:IQ:ISOLation:TIME?
Example
:CAL:IQ:ISOL:TIME?
Notes
This returns 6 integer values: year, month, day, hour, minute, second. When no calibration has
been performed, all values will be 0.
Initial S/W Revision
A.02.00
666
Remote Language Compatibility Measurement Application Reference