Marker count – Agilent Technologies Signal Analyzer N9030a User Manual
Page 722
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker
the Y-axis result. Additional information is shown for markers which have marker functions
turned on.
Turning the Marker Table on turns the Peak Table off and vice versa.
Key Path
Marker
Remote Command
:CALCulate:MARKer:TABLe[:STATe] OFF|ON|0|1
:CALCulate:MARKer:TABLe[:STATe]?
Example
CALC:MARK:TABL ON turns on the marker table.
Preset
OFF
State Saved
The on/off state of the Marker Table is saved in instrument state
Initial S/W Revision
Prior to A.02.00
Marker Count
Accesses the marker count menu.
Key Path
Marker
Readback line
[On] if count on for the selected marker, [Off] if it is off.
Initial S/W Revision
Prior to A.02.00
Counter
Turns the marker frequency counter on and off. The selected marker is counted, and if the
selected marker is a delta marker and its reference marker is not fixed, the reference marker is
counted as well.
See
"Understanding the Marker Counter" on page 724
.
See
"Query Count Value" on page 723
Key Path
Marker, Marker Count
Remote Command
:CALCulate:MARKer[1]|2|...12:FCOunt[:STATe] OFF|ON|0|1
:CALCulate:MARKer[1]|2|...12:FCOunt[:STATe]?
Example
CALC:MARK2:FCO ON selects marker 2, turns it on, and turns on the counter
CALC:MARK2:FCO:X? returns the counted frequency.
722
Remote Language Compatibility Measurement Application Reference