3 Programming the Analyzer
STATus Subsystem
Operation Positive Transition
This command determines which bits in the Operation Condition register will set the
corresponding bit in the Operation Event register when the condition register bit has a positive
transition (0 to 1). The variable
want to enable.
Mode
All
Remote Command
:STATus:OPERation:PTRansition
:STATus:OPERation:PTRansition?
Example
STAT:OPER:PTR 1 Align Now operation beginning will be reported to the Status Byte Register.
Preset
32767
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Preset the Status Byte
Sets bits in most of the enable and transition registers to their default state.
Presets all the Transition Filters, Enable Registers, and the Error/Event Queue Enable.
Has no effect on Event Registers, Error/Event Queue, IEEE 488.2 ESE, and SRE Registers as
described in
.
Remote Command
:STATus:PRESet
Example
STAT:PRES
Initial S/W Revision
Prior to A.02.00
Questionable Register
The following commands and queries are available for this register:
"Questionable Condition " on page 144
"Questionable Enable " on page 144
"Questionable Event Query " on page 145
Remote Language Compatibility Measurement Application Reference