Agilent Technologies Signal Analyzer N9030a User Manual
Page 801
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
Preset
Limits 1 and 2 preset to 1, Limits 3 and 4 preset to 2, Limits 5 and 6 preset to 3
Not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved
Saved in instrument state.
Min
1
Max
6
Readback
Trace 1|2|3|4|5|6
Initial S/W Revision
A.02.00
Test Trace
Selects the trace you want the limit to test. A limit is applied to one and only one trace; each trace
can have both an upper and a lower limit. When executing Limit Test, the limit is applied only to
the specified trace.
A trace can have multiple limit lines simultaneously; in that case, only one upper and one lower
limit line will affect the color of the trace. Other limit lines will be displayed, and will affect the
pass/fail status, but the trace will not turn red if it crosses a secondary limit line.
Key Path
Meas Setup, Limits, Properties
Remote Command
:CALCulate:LLINe[1]|2|...6:TRACe 1|2|3|4|5|6
:CALCulate:LLINe[1]|2|...6:TRACe?
Example
:CALC:LLIN3:TRAC 2 applies limit 3 to trace 2.
Notes
When the trace display is off, the trace is not tested. The trace is tested only when the trace
display is on and Test Limits (see
) is on.
Couplings
This matters when testing a trace or limit line for failure, via :CALC:LLIN3:FAIL? or
:CALC:TRAC2:FAIL?
Preset
Limits 1 and 2 preset to 1, Limits 3 and 4 preset to 2, Limits 5 and 6 preset to 3
Not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved
Saved in instrument state.
Min
1
Max
6
Readback
Trace 1|2|3|4|5|6
Initial S/W Revision
A.02.00
Remote Language Compatibility Measurement Application Reference
801