6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
The table editor will only operate properly if the analyzer is sweeping, because its updates are tied to the
sweep system. Thus, you should not try to use the editor in single sweep, and it will be sluggish during
compute-intensive operations like narrow-span FFT sweeps.
When exiting the edit menu (by using the Return key or by pressing an instrument front panel
key), the editor window turns off, however the Limit is still on and displayed,
and the
amplitude scale remains Log.
Limits are turned off by a Preset, but the Limits arrays (data) are only reset (deleted) by Restore
Mode Defaults. They survive shutdown and restarting of the analyzer application, which means
they will survive a power cycle.
Key Path
Meas Setup, Limits
Couplings
A remote user can enter or access limit line data via :CALCulate:LLINe[1]|2|3|4|5|6:DATA
Initial S/W Revision
A.02.00
Navigate
Lets you move through the table to edit the desired point
Key Path
Meas Setup, Limits, Edit
Notes
There is no value readback on the key
Min
1
Max
2000
Initial S/W Revision
A.02.00
Frequency
Lets you edit the frequency of the current row.
Key Path
Meas Setup, Limits, Edit
Notes
There is no value readback on the key
Min
0
Max
1 THz
Initial S/W Revision
A.02.00
Remote Language Compatibility Measurement Application Reference