6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
OUTPut:ANALog:AUTO?
Example
OUTP:ANAL:AUTO ON
Preset
ON
State Saved
Saved in Input/Output State
Initial S/W Revision
A.04.00
Off
Turns off the analog output.
Key Path
Input/Output, Output Config, Analog Out
Example
OUTP:ANAL OFF ! causes the analog output to be off
Readback Text
Off
Initial S/W Revision
A.04.00
Screen Video
Selects the analog output to be the screen video signal. In this mode, the pre-detector data is
output to the Analog Out connector. The output looks very much like the trace displayed on the
analyzer’s screen, and depends on the Log/Lin display Scale, Reference Level, and dB per
division, but is not influenced by the selected detector or any digital flatness corrections or trace
post-processing (like Trace Averaging).
Note that this mode is similar to the Analog Output of the HP 8566 family and the Video Out (opt
124) capability of the Agilent PSA analyzer (E444x), although there are differences in the
behavior.
Key Path
Input/Output, Output Config, Analog Out
Example
OUTP:ANAL SVID
Dependencies
Because the Screen Video output uses one of the two IF processing channels, only one detector
is available while Screen Video is selected. All active traces will change to use the same detector
as the selected trace when Screen Video is activated.
Screen Video output is not available while any EMI Detector is selected (Quasi Peak, RMS
Average or EMI Average), because these detectors use both IF processing channels.
Consequently, if the user chooses an EMI Detector, there will be no Screen Video output.
The output holds at its last value during an alignment and during a marker count. After a sweep:
•
If a new sweep is to follow (as in Continuous sweep mode), the output holds at its last value
during the retrace before the next sweep starts. If the analyzer is in zero-span, there is no
Remote Language Compatibility Measurement Application Reference