Agilent Technologies Signal Analyzer N9030a User Manual
Page 906

6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Peak Search
More Information
If the Display Line (see the Section “View/Display”) is turned on, the Peak Table can be selected
to include all peaks, only those above the Display Line, or only those below the Display Line. See
Figures 1–2 and 1–3 to understand what happens if both Display Line and Pk Threshold are
turned on.
Figure 1- 2 Above Display Line Peak Identification
Figure 1- 3Below Display Line Peak Identification
All
Sets the peak table to display the 20 highest peaks in the order specified by the current Peak Sort
setting. If the Peak Criteria are turned on, then only peaks that meet the defined Pk Excursion
and Pk Threshold values will be found.
Key Path
Peak Search, Peak Table, Peak Readout
Example
CALC:MARK:PEAK:TABL:READ ALL
Notes
Auto return after pressed
Readback
All
Initial S/W Revision
Prior to A.02.00
906
Remote Language Compatibility Measurement Application Reference