6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
I/O Config
Activates a menu for identifying and changing the I/O configuration for remote control.
Key Path
System
Initial S/W Revision
Prior to A.02.00
GPIB
Activates a menu for configuring the GPIB I/O port.
Key Path
System, I/O Config
Initial S/W Revision
A.02.00
GPIB Address
Select the GPIB remote address.
Key Path
System, I/O Config, GPIB
Mode
All
Remote Command
:SYSTem:COMMunicate:GPIB[1][:SELF]:ADDRess
:SYSTem:COMMunicate:GPIB[1][:SELF]:ADDRess?
Example
:SYST:COMM:GPIB:ADDR 17
Notes
Changing the Address on the GPIB port requires all further communication to use the new
address.
Preset
This is unaffected by Preset but is set to 18 on a “Restore System Defaults->Misc”
State Saved
No
Range
0 to 30
Min
0
Max
30
Initial S/W Revision
Prior to A.02.00
GPIB Controller
Sets the GPIB port into controller or device mode. In the normal state, GPIB controller is disabled,
which allows the analyzer to be controlled by a remote computer. When GPIB Controller is
1294
Remote Language Compatibility Measurement Application Reference