6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trace/Detector
All Traces Auto
This is designed to quickly return the selected set of detectors to the “preset” state, which is auto-
selected.
Key Path
Trace/Detector, Detector, Preset Detectors
Couplings
Sets all traces’ Detector Auto to true.
Initial S/W Revision
Prior to A.02.00
Peak / Average / NPeak
This is a setting for making a measurement of the average power and the signal envelope.
Key Path
Trace/Detector, Detector, Preset Detectors
Couplings
Trace 1: Set to peak detection, and Clear-Write.
Trace 2: Set to average detection, and Clear-Write.
Trace 3: Set to negative peak detection, and Clear-Write.
Initial S/W Revision
Prior to A.02.00
Peak / Sample / NPeak
This is a setting for making a measurement that displays a power sample and the signal envelope.
Key Path
Trace/Detector, Detector, Preset Detectors
Couplings
Trace 1: Set to peak detection, and Clear-Write.
Trace 2: Set to sample detection, and Clear-Write.
Trace 3: Set to negative peak detection, and Clear-Write.
Initial S/W Revision
Prior to A.02.00
Clear Trace
Clears the selected trace (from the front panel) or the specified trace (from SCPI). Does not affect
the state of any function or variable in the instrument. Loads mintracevalue into all of the points
in the selected trace, unless the trace is in Min Hold in which case it loads maxtracevalue. It does
this even if Update=Off.
Remote Language Compatibility Measurement Application Reference