6 RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale
Example
POW:ATT:STEP 2
Notes
Note this feature works like a 1-N choice from the front panel, but it takes a specific value
(in dB) when used remotely. The only valid values are 2 and 10.
Dependencies
Blanked in CXA and EXA if option FSA (2 dB steps) is not present. If blanked, attempts to set it
via SCPI will yield an error.
Couplings
When the attenuation step size changes, the current mechanical attenuation value is adjusted
(if necessary) to be quantized to the new step size. That is, if step is set to 10 dB, mech atten is
increased if necessary so it is a multiple of 10 dB
Preset
PXA and MXA: 2 dB
EXA and CXA: 10 dB (2 dB with option FSA)
State Saved
Saved in instrument state
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.03.00
Max Mixer Level
Controls the limitation on the Ref Level for a given attenuation setting, and therefore also
interacts with the Auto rules for selecting the attenuation as a coupling from the reference level.
Key Path
AMPTD Y Scale, Attenuation
Remote Command
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?
Example
POW:MIX:RANG –15 dBm
Preset
–10 dBm
State Saved
Saved in instrument state
Min
–50 dBm
Max
–10 dBm
Default Unit
Depends on the current selected Y axis unit, see Swept SA discussion of Y Axis Unit
Initial S/W Revision
Prior to A.02.00
528
Remote Language Compatibility Measurement Application Reference