Agilent Technologies Signal Analyzer N9030a User Manual
Page 1292

6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
If the value of the Timebase DAC changes (by switching to Calibrated from User with User set to a
different value, or in User with a new value entered) an alignment may be necessary. The
alignment system will take appropriate action; which will either invoke an alignment or cause an
Alert.
Key Path
System, Alignments
Mode
All
Remote Command
:CALibration:FREQuency:REFerence:MODE CALibrated|USER
:CALibration:FREQuency:REFerence:MODE?
Example
:CAL:FREQ:REF:MODE CAL
Notes
If the value of the timebase is changed the alignment system automatically performs an
alignment or alerts that an alignment is due.
If the value of the timebase is changed the alignment system automatically performs an
alignment or alerts that an alignment is due.
Preset
This is unaffected by Preset but is set to CALibrated on a “Restore System Defaults->Align”.
State Saved
No
Initial S/W Revision
Prior to A.02.00
Calibrated
Sets the Timebase DAC to the value established during factory or field calibration. The value
displayed on the menu key is the calibrated value.
Key Path
System, Alignments, Timebase DAC
Mode
All
Example
:CAL:FREQ:REF:MODE CAL
Readback Text
[xxx]
< where xxx is the calibrated value
Initial S/W Revision
Prior to A.02.00
User
Allows setting the Timebase DAC to a value other than the value established during the factory or
field calibration. The value displayed on the menu key is the calibrated value.
Key Path
System, Alignments, Timebase DAC
1292
Remote Language Compatibility Measurement Application Reference