Agilent Technologies Signal Analyzer N9030a User Manual
Page 526
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale
COMBined
[:SENSe]:POWer[:RF]:RANGe:OPTimize:ATTenuation?
Notes
The SCPI parameter ELECtrical sets this function to On in single attenuator models. The SCPI
parameter COMBined is mapped to ELECtrical in single attenuator models; if you send
COMBined, it sets the function to On and returns ELEC to a query.
Dependencies
This key only appears in Dual Attenuator models with an Electronic Attenuator installed. It does
not appear in models with the Single Attenuator configuration, as in the single attenuator
configuration there is no “electronic attenuator” there is only a single integrated attenuator
(which has both a mechanical and electronic stage).
When Enable Elec Atten is off or grayed out, the Pre-Adjust for Min Clip key is grayed out.
Preset
OFF for Swept SA measurement; ON for all other measurements that support Pre-Adjust for Min
Clip
State Saved
Saved in instrument state
Range
Dual attenuator models:
Off | Elec Atten Only | Mech + Elec Atten
Single attenuator models:
Off | On
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.03.00
Remote Command
[:SENSe]:POWer[:RF]:RANGe:AUTO ON|OFF|1|0
[:SENSe]:POWer[:RF]:RANGe:AUTO?
Notes
ON aliases to "Elec Atten Only" (:POW:RANG:OPT:ATT ELEC)
OFF aliases to "Off" (:POW:RANG:OPT:ATT OFF)
The query :POW:RANG:AUTO? returns true if :POW:RANG:OPT:ATT is not "Off"
Initial S/W Revision
Prior to A.02.00
Off
Turns Pre-Adjust for Min Clip off. This is the default setting.
This key is grayed out in measurements that do not support this functionality. The spectrum
analyzer measurement, Swept SA, does not support this functionality.
Key Path
AMPTD Y Scale, Attenuation, Pre-Adjust for Min Clip
526
Remote Language Compatibility Measurement Application Reference