Agilent Technologies Signal Analyzer N9030a User Manual
Page 640
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
If an external frequency reference is being used, you must enter the frequency of the external
reference if it is not exactly 10 MHz. The External Ref Freq key is provided for this purpose.
Key Path
Input/Output
Remote Command
[:SENSe]:ROSCillator:SOURce:TYPE INTernal|EXTernal|SENSe
[:SENSe]:ROSCillator:SOURce:TYPE?
Preset
This is unaffected by a Preset but is set to SENSe on a "Restore Input/Output Defaults" or
"Restore System Defaults->All".
State Saved
Saved in instrument state.
Status Bits/OPC
dependencies
STATus:QUEStionable:FREQuency bit 2 set if unlocked.
Backwards Compatibility
Notes
Freq Ref In was not saved in state in the legacy instruments. It is a part of state in the X-Series.
Initial S/W Revision
Prior to A.02.00
Remote Command
[:SENSe]:ROSCillator:SOURce?
Notes
The query [SENSe]:ROSCillator:SOURce? returns the current switch setting. This means:
1. If it was set to SENSe but there is no external reference so the instrument is actually using
the internal reference, then this query returns INTernal and not SENSe.
2. If it was set to SENSe and there is an external reference present, the query returns EXTernal
and not SENSe.
3. If it was set to EXTernal, then the query returns "EXTernal"
4. If it was set to INTernal, then the query returns “INTernal”
Preset
SENSe
Backwards Compatibility
Notes
The query [:SENSe]:ROSCillator:SOURce? was a query-only command in ESA which always
returned whichever reference the instrument was using. The instrument automatically switched
to the ext ref if it was present.
In PSA (which had no sensing) the command [:SENSe]:ROSCillator:SOURce set the reference
(INT or EXT), so again its query returned the actual routing.
Thus the query form of this command is 100% backwards compatible with both instruments.
Initial S/W Revision
Prior to A.02.00
Remote Command
[:SENSe]:ROSCillator:SOURce INTernal|EXTernal
Notes
For PSA compatibility the command form is provided and is directly mapped to
[:SENSe]:ROSCillator:SOURce:TYPE
Initial S/W Revision
Prior to A.02.00
640
Remote Language Compatibility Measurement Application Reference