S2 [two] (single sweep), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 457: Description, Syntax legacy products
![background image](/manuals/40104/457/background.png)
5 Legacy Command Descriptions
S2 [two] (Single Sweep)
S2 [two] (Single Sweep)
Syntax
Legacy Products
8566A/B, 8568A/B
Description
Sets the instrument to single sweep mode. Each subsequent time that this command is sent, one
sweep is started if the trigger conditions are met.
Format
S2
Query Data Type
N/A
SCPI Equivalent Commands
:INITiate:CONTinuous 0 (see
"Cont (Continuous Measurement/Sweep)" on page 563
Notes
The functions of S2 are similar to
"SNGLS (Single Sweep)" on page 463
.
Remote Language Compatibility Measurement Application Reference
457