6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
Remote Command
:CALCulate:LIMit[1]|2|...6:CONTrol:POINts?
Example
:CALC:LIM:CONT:POIN?
returns the number of points in the limit line.
Preset
Limit line data is cleared by Restore Mode Defaults.
State Saved
Saved in instrument state.
Initial S/W Revision
A.02.00
Limit Line Upper / Lower (Remote Command Only, SCPI standard conformance)
Defines a list of amplitude values for a given limit line. Changing the number of elements in the
list spectrum will automatically turn the limit line off. Using the “UPP” syntax defines an upper
limit line, using the “LOW” syntax defines a lower limit line. Note that a line may not be
simultaneously both upper and lower; the type of the limit line will automatically be changed as
appropriate. Up to 200 points may be defined for each limit using the following parameters.
Range: –200 dBm to +100 dBm
The points query returns the number of points in the amplitude list. It will not be possible to turn
on the limit line unless the number of points in the control matches the number of points in the
amplitude.
The points query returns the number of points in the amplitude list. It should match the number
of points in the control, that is, the number of values for the CONTrol axis and for the
corresponding UPPer and/or LOWer limit lines must be identical. If one array is larger than the
other, the limit trace is built using only as much data as is contained in the smaller array.
An empty array returns the system error message “list is empty” to a data query, 0 to a POINts
query.
Remote Command
:CALCulate:LIMit[1]|2|...6:UPPer[:DATA]
:CALCulate:LIMit[1]|2|...6:UPPer[:DATA]?
Example
:CALC:LIM:UPP –10, –10, –20, –20
describes the amplitude values of an upper limit line
Preset
Limit line data is cleared by Restore Mode Defaults.
State Saved
Saved in instrument state
Initial S/W Revision
A.02.00