6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Save
Key Path
Save, Data, Amplitude Correction
Preset
Not part of a Preset, but is reset to Correction 1 by Restore Input/Output Defaults. Survives a
shutdown.
Readback
1
Initial S/W Revision
A.02.00
Amplitude Correction
These keys let you choose which Correction to save. Once selected, the key returns back to the
Export Data menu and the selected Correction number is annotated on the key.
The next step in the Save process is to select the Save As key in the Export Data menu.
Key Path
Save, Data, Amplitude Correction
Preset
Not part of a Preset, but is reset to Correction 1 by Restore Input/Output Defaults. Survives a
shutdown.
Readback
1
Initial S/W Revision
A.02.00
Trace
Pressing this key selects Traces as the data type to be exported. Pressing this key when it is
already selected brings up the Trace Menu, which allows you to select which Trace to save.
The trace file contains “meta” data which describes the current state of the analyzer. The
metadata is detailed in
"Trace File Contents" on page 1045
below.
Key Path
Save, Data
Remote Command
:MMEMory:STORe:TRACe:DATA TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 |
TRACE6 | ALL,
Example
:MMEM:STOR:TRAC:DATA TRACE2, "myTrace2.csv" Exports the 2nd trace to the file
myTrace2.csv in the current path. The default path is My Documents\SA\data\traces
Notes
If the save is initiated via SCPI, and the file already exists, the file will be overwritten.
Using the C: drive is strongly discouraged, since it runs the risk of being overwritten during an
instrument software upgrade.
Both single and double quotes are supported for any filename parameter over SCPI.
Dependencies
For SA measurements, traces cannot be recalled from a trace file that was saved with ALL traces
selected.
1044