Agilent Technologies Signal Analyzer N9030a User Manual
Page 1404
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
Remote Command
:TRIGger[:SEQuence]:VIDeo:DELay
:TRIGger[:SEQuence]:VIDeo:DELay?
:TRIGger[:SEQuence]:VIDeo:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:VIDeo:DELay:STATe?
Example
TRIG:VID:DEL:STAT ON
TRIG:VID:DEL 100 ms
Notes
Video trigger delay may be set to negative values, in time domain, FFT and even swept. It makes
intuitive sense in time domain and works well in FFT mode where the bandwidth of the filter
before the video trigger is about 1.25 span. In swept spans, negative settings of Trig Delay are
treated as a zero setting within the internal hardware and the advisory message "Neg. Trig Delay
unavailable in Swept Mode, zero delay used." is generated when such a delay is set.
Preset
Off, 1 us
State Saved
Saved in instrument state
Min
–150 ms
Max
+500 ms
Default Unit
s
Backwards Compatibility
Notes
! For backward compatibility with VSA/PSA comms apps
:TRIGger[:SEQuence]:IF:DELay
:TRIGger[:SEQuence]:DELay
The legacy :TRIGger[:SEQuence]:DELay command affects the delay for the VID, LINE, EXT1,
EXT2, and RFB triggers.
Initial S/W Revision
Prior to A.02.00
Remote Command
:TRIGger[:SEQuence]:DELay
:TRIGger[:SEQuence]:DELay?
:TRIGger[:SEQuence]:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:DELay:STATe?
Example
TRIG:DEL 1 ms
Preset
1 us
State Saved
Saved in instrument state
Backwards Compatibility
Notes
In ESA/PSA, the Trigger Delay was global to all triggers. In the X-Series, the delay can be set
individually for each Trigger Source. For backward compatibility, the global DELay command
updates all instances of trigger slope (VID, LINE, EXT1, EXT2) except TV and RFBurst. The query
returns the trigger delay setting of the currently selected trigger source.
Initial S/W Revision
Prior to A.02.00
1404
Remote Language Compatibility Measurement Application Reference