5 Legacy Command Descriptions
ACPALTCH (Adjacent Channel Power Alternate Channels)
ACPALTCH (Adjacent Channel Power Alternate Channels)
Syntax
Legacy Products
8560 series
Description
Sets the number of alternate channels to be measured by an adjacent channel power measurement
to either 0, 1, or 2. The number of alternate channels is used with the command
(Adjacent Channel Power Measurement Results)" on page 226
.
Specifying parameter value 0 makes the measurement with the adjacent channel pair, but no
alternate channels. Specifying 1 selects the first alternate channel pair, which is centered at ±2
times the channel spacing away from the center frequency of the main channel. Specifying 2
selects the second alternate pair, which is at ±3 times the channel spacing.
Format
ACPALTCH
ACPALTCH?
Query Data Type
SCPI Equivalent Commands
None
Preset
Default: 0. Not affected by preset or Power Cycle.