6  RLC Swept SA Measurement Front-Panel & SCPI Reference

System

No confirmation is required when Alert is configured through a remote command.

Key Path

System, Alignments, Auto Align, Alert

Mode

All

Example

:CAL:AUTO:ALER WEEK

Readback Text

7 days

Status Bits/OPC
dependencies

Bit 14 is set in the Status Questionable Calibration register.

Initial S/W Revision

Prior to A.02.00

None

With Auto Align Alert set to None the instrument will not signal an alert. This is provided for rare
occasions where you are making a long measurement which cannot tolerate Auto Align
interruptions, and must have the ability to capture a screen image at the end of the measurement
without an alert posted to the display. Agilent does not recommends using this selection in any
other circumstances, because of the risk of accuracy performance drifting well beyond expected
levels without the operator being informed.

For front panel operation, confirmation is required to transition into this setting of Alert. The
confirmation dialog is:

Remote Language Compatibility Measurement Application Reference