Agilent Technologies Signal Analyzer N9030a User Manual
Page 1384
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trace/Detector
Example
TRAC3:DISP 1 Shows the reference trace.
Notes
Use the TRAC3:DISP command to show or blank the reference trace
Trace 3 is always the reference trace by definition.
State Saved
Saved in instrument state.
Initial S/W Revision
Prior to A.02.00
Norm Ref Lvl
Sets the level (in dB) of the normalized reference.
Key Path
Trace/Detector, Normalize
Remote Command
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:NRLevel
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:NRLevel?
Example
DISP:WIND:TRAC:Y:NRL .10 dB
DISP:WIND:TRAC:Y:NRL?
Preset
0 dB
State Saved
Saved in instrument state.
Min
–327.6 dB
Max
327.6 dB
Initial S/W Revision
Prior to A.02.00
Open/Short Cal
Performs a guided open/short calibration, while providing step-by-step instructions to the user.
This is the most accurate way to make the return loss measurement on the X-series analyzers. You
are directed through a 1-Port coaxial open calibration, and a 1-Port coaxial short calibration. The
result can then be saved to Trace 3. It is used to perform calibrated scalar reflection
measurements (return loss), using the Normalize function.
Key Path
Trace/Detector, Normalize
Mode
SA
Notes
Does not auto return
Dependencies
Key is grayed out unless Source Mode is Tracking, and control returns to the Normalize menu.
Initial S/W Revision
A.06.01
1384
Remote Language Compatibility Measurement Application Reference