Agilent Technologies Signal Analyzer N9030a User Manual
Page 825
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
Limit Line Fail? (Remote Command Only)
Tests a limit line against its associated trace. Returns a 0 if the trace is within the limit and
margin, a 1 if the trace exceeds either the limit or the margin.
Note that this command only tests one limit line – other limit lines are not tested when executing
this command. To see whether a trace passed all limits, use :CALCulate:TRACe:FAIL?.
Note this command performs the test regardless of whether the trace or the limit is turned on the
display.
Remote Command
:CALCulate:LLINe[1]|2|...6:FAIL?
Example
:CALC:LLIN:FAIL?
returns a zero if limit line 1’s associated trace has no failure, 1 if there is a margin or limit failure.
Initial S/W Revision
A.02.00
Limit Line Control (Remote Command Only, SCPI standard conformance)
Defines a list of limit line control (frequency or time) values for a given limit line. Up to 2000
points may be defined for each limit using the following parameters.
Hz (for frequency) and seconds (for time).
Range: –30 Gs to +30 Gs for time limits, –3 kHz to +1200 GHz for frequency limits.
Note that X values may be repeated if a vertical step in the limit line is desired.
The points query returns the number of points in the control. It should match the number of
points in the amplitude, that is, the number of values for the CONTrol axis and for the
corresponding UPPer and/or LOWer limit lines must be identical. If one array is larger than the
other, the limit trace is built using only as much data as is contained in the smaller array.
An empty array returns not a number (9.91e+37 to a data query), 0 to a POINts query.
Remote Command
:CALCulate:LIMit[1]|2|...6:CONTrol[:DATA]
:CALCulate:LIMit[1]|2|...6:CONTrol[:DATA]?
Example
:CALC:LIM:CONT 1GHz, 2GHz, 2GHz, 3GHz
describes the X values of a stair-stepped limit line.
Preset
Limit line data is cleared by Restore Mode Defaults.
State Saved
Saved in instrument state
Initial S/W Revision
A.02.00
Remote Language Compatibility Measurement Application Reference
825