6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup

"All other Models" on page 835

Key Path

Meas Setup, PhNoise Opt

Remote Command

[:SENSe]:FREQuency:SYNThesis:AUTO[:STATe] OFF|ON|0|1

[:SENSe]:FREQuency:SYNThesis:AUTO[:STATe]?

Example

FREQ:SYNT:AUTO ON

Preset

ON

State Saved

Saved in instrument state.

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.12.00

Models with Option EP1

Auto will choose:

Fast Tuning whenever:

Span > 44.44 MHz, or when

RBW > 1.9 MHz, or if

Source Mode is set to “Tracking”

otherwise Auto will choose Best Close in Phase Noise whenever:

Center frequency is < 195 kHz, or when

CF >= 1 MHz and Span <= 1.3 MHz and RBW <= 75 kHz

otherwise, Auto will choose Best Wide-offset Phase Noise

The RBW to be used in the calculations above is the equivalent –3 dB bandwidth of the current
RBW filter.

These rules apply whether in swept spans, zero span, or FFT spans.

Models with Option EP2

Auto will choose:

Best Close-in Φ Noise whenever:

CF < 130 kHz , or when

CF > 12 MHz  and Span < 495 kHz and RBW < 40 kHz

Otherwise, Auto will choose Fast Tuning whenever: