6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
the range to either the high or low gain setting. But you must know that your measurement
conditions will not overload the IF (in the high gain range) and that your signals are well above
the noise floor (for the low gain range), and that the signals are not changing.
Key Path
Meas Setup
Remote Command
[:SENSe]:IF:GAIN:FFT[:STATe] AUTOrange|LOW|HIGH
[:SENSe]:IF:GAIN:FFT[:STATe]?
Couplings
As with most parameters with an AUTO state, AUTO COUPLE sets it to Auto, which then picks
AUTOrange, and setting any specific value (AUTOrange, LOW or HIGH) will set the AUTO state
to false.
Preset
AUTOrange
State Saved
Saved in instrument state.
Readback Line
Autorange, High Gain or Low Gain
Initial S/W Revision
Prior to A.02.00
Auto
Allows the instrument to pick the FFT IF Gain method as appropriate. This “Auto” state is set by
the Auto Couple key, and it puts it in Autorange.
Key Path
Meas Setup
Remote Command
[:SENSe]:IF:GAIN:FFT:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:IF:GAIN:FFT:AUTO[:STATe]?
Example
IF:GAIN:FFT:AUTO ON
Preset
ON
State Saved
Saved in instrument state.
Backwards
Compatibility SCPI
DISPlay:WINDow[1]:TRACe:Y[:SCALe]:LOG:RANGe:AUTO
Included for ESA compatibility
Backwards
Compatibility SCPI
[:SENSe]:ADC:RANGe AUTO|NONE
Included for PSA compatibility. Accepted without error but ignored; the query is ignored as well
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference