Agilent Technologies Signal Analyzer N9030a User Manual
Page 660
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
retrace (after a sweep and before the next sweep starts).
This function depends on optional capability; the key will be blanked and the command will
generate an “Option not available” error unless you have Option YAV licensed in your
instrument.
Couplings
Linear Video output changes while in FFT Sweeps, so for measurements that use exclusively FFT
Sweeps, or if the user manually chooses FFT Sweeps, the Linear Video output will look different
than it does in swept mode.
Readback Text
Linear Video
Initial S/W Revision
A.04.00
Demod Audio
Selects the analog output to be the demodulation of the video signal.
When Demod Audio is selected, the demodulated audio signal appears at this output whenever
the Analog Demod application is demodulating a signal or when Analog Demod Tune and Listen is
operating in the Swept SA measurement.
When Analog Out is in the Auto state, this output is auto-selected when in the Analog Demod
mode or when Analog Demod Tune and Listen is operating in the Swept SA measurement.
If any other Analog Output is manually selected when in the Analog Demod mode or when Analog
Demod Tune and Listen
is operating in the Swept SA measurement, a condition warning message
appears.
Key Path
Input/Output, Output Config, Analog Out
Example
OUTP:ANAL DAUD
Dependencies
This key only appears if the Analog Demod application (N9063A), the N6141A or W6141A
application, or Option EMC is installed and licensed, otherwise the key will be blanked and the
command will generate an “Option not available” error.
The output holds at its last value during an alignment and during a marker count. It is not held
between sweeps, in order for Tune and Listen to work properly.
When Demod Audio is the selected Analog Output:
•
all active traces are forced to use the same detector.
•
CISPR detectors (QPD, EMI Avg, RMS Avg) are unavailable
Readback Text
Demod Audio
Initial S/W Revision
Prior to A.02.00 (this was the default functionality, and there was no selection)
Modified at S/W
Revision
A.04.00