beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 98

background image

4.2 Lead determination with standard addition

746 VA Trace Analyzer / 747 VA Stand

4-13

dU.front/dU.rear

Determination of base points: automatic

S.front/S.rear

Determination of the slopes at the base points:
automatic

I.scale

Scaling of the current axis: automatic

U.div

Division of the voltage axis for printout

Quantity

Evaluation quantity: peak height

Press the softkey

< >> >

. The subpage of "SUBSTANCES" appears with

the parameters for calibration of the substance lead:

Method: TESTPb SUBSTANCES

Pb - SegPb

Calibration 1997-03-20 16:23 Coefficients

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДД

Technique std.add Y.reg -1.957e-07

Curve type linear Slope -3.907e-05

Nonlin.

Mean dev. 2.137e-09

Additions

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД

Soln.name PbStd

ДДДДДДДДДД

ДДДДДДДДДД

Mass conc. 1 g/L g/L g/L g/L

Range min g/L g/L g/L g/L

Range max g/L g/L g/L g/L

M.conc./cm g/L g/L g/L g/L

Modify

column

Modify

line

Select

subst.

Link to

segment

Delete

subst.

Subst.

name

>>

The dialog page "SUBSTANCES" contains a subpage with the parameters for the
calibration. Here, the type of calibration (standard addition, calibration curve or internal
standard) and the required solutions are defined, in addition the results of the last
calibration are displayed.
You will find further information regarding this dialog page in section 5.6.3.

The following is a brief explanation of the most important parameters:

Technique

Calibration technique: standard addition

Curve type

Curve type for calculation of the standard addition:
linear

Soln.name

Name of the standard addition solution

PbStd

called up in the instruction

ADD>M

Mass conc.

Mass concentration of the standard addition solution:
1 g/L

4

Switch to dialog page "CALCULATION"

Press the yellow key. The dialog page

"CALCULATION" appears for calculation of the final results: