5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-301
TPURGE
Page:
O, S
Stand:
M, A
Abbreviation:
T
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. TPURGE .....
Function
Purging for a preset line time.
Parameter
Range; Default value / Meaning
t/s
0.1...999.9 s; -
Purging time
The duration of purging is entered in the line time.
Description
The
TPURGE
instruction switches on the inert gas purging for the
measuring vessel on the 747 VA Stand for the period preset in
t/s
.
Purging is used to remove the electrochemically active and hence
interfering oxygen in aqueous solutions. With the inert gas flow rate of
ca. 20 L/h set on the 747 VA Stand, a purging time of ca. 3...5 min
usually suffices.
Comments
•
If the purging should not be switched off automatically after a definite
time, the
PURGE
instruction can also be used.
•
To achieve an effective purging of the analysis solution as quickly as
possible, the solution should also be stirred with the
STIR
instruction.
see also
PURGE, ØPURGE, STIR, TSTIR