301 tpurge, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual

Page 452

background image

5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-301

TPURGE

Page:

O, S

Stand:

M, A

Abbreviation:

T

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. TPURGE .....

Function

Purging for a preset line time.

Parameter

Range; Default value / Meaning

t/s

0.1...999.9 s; -

Purging time

The duration of purging is entered in the line time.

Description

The

TPURGE

instruction switches on the inert gas purging for the

measuring vessel on the 747 VA Stand for the period preset in

t/s

.

Purging is used to remove the electrochemically active and hence
interfering oxygen in aqueous solutions. With the inert gas flow rate of
ca. 20 L/h set on the 747 VA Stand, a purging time of ca. 3...5 min
usually suffices.

Comments

If the purging should not be switched off automatically after a definite
time, the

PURGE

instruction can also be used.

To achieve an effective purging of the analysis solution as quickly as
possible, the solution should also be stirred with the

STIR

instruction.

see also

PURGE, ØPURGE, STIR, TSTIR