Metrohm 746 VA Trace Analyzer User Manual

Page 136

background image

4.4 Lead determination with internal standard

746 VA Trace Analyzer / 747 VA Stand

4-51

The full report

Full report

contains not only the usual header (see

section 4.2.6), but also the following important sections:

1

Data for substance

Only the substance name

Pb is given, the other

evaluation of Pb

parameters remain blank as they are not calcula-
ted when the calibration curve is recorded.

2

Data for substance

Only the substance name

CdRef is given, the

evaluation of Cd

other parameters remain blank as they are not
calculated when the calibration curve is recorded.

3

Data for calibration

Only the proportionality factor is given.

4

Final results

The final result is not calculated as the variable
MC:Pb is not calculated when the calibration
curve is recorded.

Finally, the calibration curve is printed out:

================== METROHM 746 VA TRACE ANALYZER (5.746.0101) ==================

Method : INTSTD User : Date : 1996-12-16

Mode : DPMODE Determ. : 12161422 Time : 14:22:33

Run : 4 Sample :

Internal standard calibration curve

Curve type : 0/lin Slope : 5.699e-01 Mean.dev. : 3.553e-02

Substance : Pb

The calibration curve contains the usual header (see section 4.2.6) and

the following important sections:

1

Data for calibration curve Curve type, proportionality factor Slope, mean

deviation of the measured values about the
calibration curve.

2

Calibration curve

Measured points (mean values of the measured
peak currents for the calibration solutions),
calculated calibration curve, current axis, axis for
effective mass concentrations of analyte and
reference substance in the measuring vessel.

2