5.6.3 SUBSTANCES

746 VA Trace Analyzer / 747 VA Stand

5-141

Method: XXXXXXXX SUBSTANCES

........ - XXXXXXXX

Calibration XXXX-XX-XX XX:XX

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД

Technique none

Modify

column

Modify

line

Select

subst.

Link to

segment

Delete

subst.

Subst.

name

>>

Softkeys

Meaning

Meaning of the softkeys F1...F7, see page 130

>>

Switching to subpage "Substance parameters"

Switching to the subpage "Substance parameters" with the parame-
ters for substance recognition, baseline calculation, scaling and
evaluation.

Parameter

Range; Default value / Meaning

Technique

none, std.add., cal.crv., int.std.; none

Calibration technique

In order to determine the sample content (mass concentration) from
the measured current or area values for the substances

EV:

,

a calibration is required (see section 6.9). One of the following calibra-
tion techniques can be selected for this (with

none

, no calculation is

performed):

none

None

std.add.

Standard addition

cal.crv.

Calibration curve

int.std.

Internal standard

21

20

19

20

19

22
23
24