beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 340

background image

5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-189

The phase dependent differences in the current measurements give
curves which can be evaluated using the baselines of type

AC2 rel.

or

AC2 abs.

(see section 5.6.3).

U / 2

I.abs

I

U

Comments

The

AC2MODE

measurement mode can be used with all types of

electrodes together with the instructions

MEAS

,

SWEEP

,

DSWEEP

,

FSWEEP

or

TSWEEP

.

The electrode type

DME

,

SMDE

,

HMDE

or

RDE

must be entered before

the

AC2MODE

measurement mode.

With the

AC2MODE

measurement mode, the potentiostat always

operates in the operating mode "fast".

SWEEP

and

FSWEEP

are thus

identical for this mode.

The quantity

t.drop

important for the definition of

t.step

depends

on the type of electrode selected:

DME

t.drop = 20 ms

HMDE and RDE

t.drop = 0 ms

SMDE

t.drop = Drop.size

×× 40 ms

see also

AC1MODE, SWEEP, DME, SMDE, HMDE, RDE

Σ

(I

n+

– I

n–

)

n

I

=

I.rel

= I

f

+ I

r

I

f

I

r

n = Meas.cycles