7.7 Standard operating procedure in the context of GLP/ISO 9001 guidelines

746 VA Trace Analyzer / 747 VA Stand

7-61

7.7.4

Performing the determination

To perform the lead determination in the prepared sample solution, the following

steps are necessary:

1 Switch to dialog page ”MONITORING”

Press the yellow key.

2 Switching the program run mode to

determination

If the paramete

r Run mode

is set to

calibration

, proceed as follows

Use the cursor keys to select the field

Run mode

.

Press the key.

Press the softkey

and confirm this entry with the

key.

3 Entry of the sample volume

Enter the sample volume as follows:

Use the cursor keys to select the first field below the title

Sample size/S0

.

Press the key.

Enter

0.1 mL

and confirm this entry with the key.

4 Start the determination

Press the red key.

The sample solution is measured three times. On completion of the third

measurement, the program is interrupted. The screen shows the prompt

to add 50

µ

L of the standard addition solution

PbStd

to the measuring

vessel

ADD PbStd 50.000 uL > M

.

5 First standard addition

Remove stopper 25

25 and add 50

µ

L of the lead ion standard solution to

the measuring vessel through the pipetting opening using a pipette. Then

close the pipetting opening with stopper 25

25 and press the

key.

The spiked sample solution is measured three times. On completion of

the third measurement, the program is interrupted. The screen again

shows the prompt to add 50

µ

L of standard addition solution

PbStd

to

the measuring vessel

ADD PbStd 50.000 uL > M

.

6 Second standard addition

Remove stopper 25

25 and add 50

µ

L of the lead ion standard solution to

the measuring vessel through the pipetting opening using a pipette. Then

close the pipetting opening with stopper 25

25 and press the

key.

The sample solution spiked a second time is measured three times. On

completion of the third measurement, the results are calculated and the

reports and curves printed out (see section 4.2).