6 Data processing and evaluation

746 VA Trace Analyzer / 747 VA Stand

6-26

EV

a

0

linear

nonlin

std.add.

The non-linear curve shown in the new coordinate grid is not de-

scribed completely by the parameters a, b and c determined for the

original curve.

6

Calculation of the mass concentration

ρρ

s

s

The found, effective mass concentration

ρ

eff,s

eff,s

is converted into the sought

mass concentration

ρ

ss

by taking the dilution into account (see section 6.9.2):

Mass conc.

=

ρ

ss

=

ρ

eff,s

eff,s

/ f

VV

7

Calculation of the total scatter

σσ

x

x

As measurement of the sample is part of the calibration in standard addition,

for the determination of the total scatter

σ

xx

only the error contribution from

the calibration must be taken into account (see section 6.9.3):

σ

xx

=

σ

x,C

x,C

As the total scatter

σ

xx

refers to the effective mass concentration

ρ

eff,s

eff,s

, it must

also be converted by taking the dilution into account:

MC.dev.

=

σ

xx

/ f

VV

To be able to calculate the total scatter

MC.dev.

, with the linear curve type at

least 3, with the nonlinear curve at least 4 measured value pairs must be

available, otherwise

MC.dev.

= 0 will be set and the message

no

meas.deviation

put out on the "RESULTS/Evaluations" page.

Standard addition is the usual calibration method for the majority of the applications

possible with the 746 VA Trace Analyzer. Its advantage lies in its high dependability

as the calibration in the sample takes place under real matrix conditions and all

measurement parameters remain unchanged. With regard to optimum accuracy

and scatter, several rules must be observed in standard addition:

Check linearity range

In development of the method, the linearity range should be checked for each

substance. This involves performing several standard additions over a wide

concentration range. Using the calibration curve shown on the "CURVES"

page, you can then determine the region in which the standard addition is

linear and that in which it is nonlinear.

Working in the linear range

If the substance content lies in the linear range, multiple standard addition is

meaningful only if you wish to check the linearity in every determination. To

reduce the scatter, it is better to spike the solution once only and to choose

as many replications as possible.

ρρ

eff

eff

Sample

Standard solution