beautypg.com

7 safety, errors, troubleshooting, diagnosis, glp – Metrohm 746 VA Trace Analyzer User Manual

Page 507

background image

7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-8

Error message

Meaning/Causes

Rectification

Exit

CALCUL - ERROR 1:
Variables list overflow

Number of different variables

> 25

Reduce number of variables on the

”CALCULATION” page

Esc

CALCUL - ERROR 2:
Illegal identifier

Variable name too long

(> 8 characters)

Inadmissible assignment

(e.g. 5 = 7)

Inadmissible constant

(e.g. A = 5.4+)

Shorten variable name to max.

8 characters

Check formulae on the

”CALCULATION” page and change

Check formulae on the

”CALCULATION” page and change

Esc

CALCUL - ERROR 3:
Illegal write to system
variable

Unlawful overwriting of a

system variable

(e.g.

S0

,

S1

,

S2

,

S3

)

Check formulae on the

”CALCULATION” page and change

Esc

CALCUL - ERROR 4:
Expects token (=)

Missing equals sign

Check formulae on the

”CALCULATION” page and change

Esc

CALCUL - ERROR 5:
Illegal use before set

Variable used without previous

assignment

Check formulae on the

”CALCULATION” page and change

Esc

CALCUL - ERROR 6:
Bracket nesting level
too deep

Too many nested brackets

Check formulae on the

”CALCULATION” page and change

Esc

CALCUL - ERROR 7:
Illegal nesting of
brackets

Number of opening and closing

brackets not identical

Check brackets in formulae on the

”CALCULATION” page and change

Esc

CALCUL - ERROR 8:
System variable not
found

System variable used in the

formula has not been defined

(e.g.

S1

without entry)

Enter system variable on the

”SAMPLES” pages or modify formulae

Esc

CALCUL - ERROR 9:
Illegal multiple assign
(R)

Unlawful multiple assignment to

same variable

(e.g.

R1 = aaa

;

R1 = bbb

)

Check formulae on the

”CALCULATION” page and change

Esc

CALCUL - ERROR 10:
Unexpected end of
formula

End of formula not defined

(e.g.

R2 =

)

Check formulae on the

”CALCULATION” page and change

Esc

CALCUL - ERROR 11:
Value out of range

Constant in the formula lies

outside the admissible numeric

range (–3.7.10

33

... –1.7.10

–33

/ 0

/ 1.7.10

–33

... 3.7.10

33

)

Check formulae on the

”CALCULATION” page and change

Esc

CALCUL - ERROR 12:
Undefined error

Undefined error

Check formulae on the

”CALCULATION” page and change

Esc

CALIBR - ERROR 1:
Calibration procedure
inadequate

Unsuitable procedure in non-

linear standard addition, no

evaluation possible

Select linear standard addition

Increase amount/concentration of

standard addition solution

Increase number of standard addi-

tions

E/C

CALIBR - ERROR 2:
Illegal dilution

Inadmissible dilution (dilution

factor >1 or <10

-6

) or faulty

program flow in conjunction

with calibration (no evaluation

possible)

SMPL

instruction missing

Check dilution of sample (dispensing

instructions)

Check ”SUBSTANCES/Calibration”

page

E/C

CALIBR - ERROR 3:
Invalid result

Invalid result, mass concentra-

tion >10 kg/L (no evaluation

possible)

Check program or calibration

E/C

CALIBR - ERROR 4:
Difference of calibr.
solutions too low

Current or concentration

change in standard additions or

in plotting of calibration curves

<1%

Check standard addition or calibration

solutions

E/C

CALIBR - ERROR 5:
Missing calibration
coefficients

No coefficients for calibration

curve available

Rerecord calibration curve or enter

coefficients on the

”SUBSTANCES/Calibration” page

E/C