Metrohm 746 VA Trace Analyzer User Manual
Page 45

3 Installation
746 VA Trace Analyzer / 747 VA Stand
3-18
3.4.4
Filling the capillary without vacuum
The glass capillary 109
109 can normally be filled with mercury by the method de-
scribed here without vacuum. However, if difficulties regarding stability or repro-
ducibility arise with a capillary filled in this manner, try to fill the capillaries by the al-
ternative method with vacuum (section 3.4.5).
To fill the mounted glass capillary 109
109 (section 3.4.3) with Hg without vacuum, pro-
ceed as follows:
1 Install multi-mode electrode in 747 VA Stand
•
With the measuring head arm 40
40 tilted back, slide the empty measuring
vessel 28
28 into the holder 26
26 of the 747 VA Stand and then lower the
measuring head arm 40
40.
•
Carefully insert multi-mode electrode 38
38 in opening 76
76 of the measuring
head 41
41 (during insertion, the tip of the capillary 109
109 must not touch the
measuring head) and push in carefully as far as it will go. Avoid water
drops touch the tip of the capillary.
2 Connect multi-mode electrode
•
Screw FEP tubing 51
51 for the inert gas supply into connection 93
93 of the
multi-mode electrode 38
38.
•
Screw FEP tubing 59
59 for the inert gas supply into connection 94
94 of the
multi-mode electrode 38
38.
•
Attach electrode cable 37
37 (WE) to screw connection 110
110 of the multi-
mode electrode 38
38: push cable lug under the screw and then tighten
screw firmly.
3 Fill capillary with mercury
•
Switch on 746 VA Trace Analyzer with mains switch 11
11 (746 VA Trace
Analyzer and 747 VA Stand must first be installed properly as described
in section 3.2 and section 3.3).
•
Select dialog page "STAND" and then switch on the inert gas supply to
the 747 VA Stand by pressing the softkey
. This pressurizes
the multi-mode electrode 38
38 and the mercury begins to flow slowly out of
the capillary.
•
Gently tap the MME with your finger (to remove any air bubbles) and
allow the mercury to flow out of the capillary into the empty measuring
vessel for approx. 2 min.
•
Fill measuring vessel 28
28 with 10 mL ultrapure water and add 1 drop KCl
solution (in pure water, mercury drops from the capillary only with diffi-
culty).
•
Allow mercury to flow out of the capillary for ca. 2 min while checking the
drop formation: The drop time should be ca. 3 s.
4 Adjusting the sealing needle 97
97
•
Turn slotted screw 96
96 using a suitable coin slowly in a clockwise direction
until the mercury flow stops.
•
Open slotted screw 96
96 slightly in an anticlockwise direction until the
mercury flow restarts.