3 Installation

746 VA Trace Analyzer / 747 VA Stand

3-18

3.4.4

Filling the capillary without vacuum

The glass capillary 109

109 can normally be filled with mercury by the method de-

scribed here without vacuum. However, if difficulties regarding stability or repro-

ducibility arise with a capillary filled in this manner, try to fill the capillaries by the al-

ternative method with vacuum (section 3.4.5).
To fill the mounted glass capillary 109

109 (section 3.4.3) with Hg without vacuum, pro-

ceed as follows:

1 Install multi-mode electrode in 747 VA Stand

With the measuring head arm 40

40 tilted back, slide the empty measuring

vessel 28

28 into the holder 26

26 of the 747 VA Stand and then lower the

measuring head arm 40

40.

Carefully insert multi-mode electrode 38

38 in opening 76

76 of the measuring

head 41

41 (during insertion, the tip of the capillary 109

109 must not touch the

measuring head) and push in carefully as far as it will go. Avoid water

drops touch the tip of the capillary.

2 Connect multi-mode electrode

Screw FEP tubing 51

51 for the inert gas supply into connection 93

93 of the

multi-mode electrode 38

38.

Screw FEP tubing 59

59 for the inert gas supply into connection 94

94 of the

multi-mode electrode 38

38.

Attach electrode cable 37

37 (WE) to screw connection 110

110 of the multi-

mode electrode 38

38: push cable lug under the screw and then tighten

screw firmly.

3 Fill capillary with mercury

Switch on 746 VA Trace Analyzer with mains switch 11

11 (746 VA Trace

Analyzer and 747 VA Stand must first be installed properly as described

in section 3.2 and section 3.3).

Select dialog page "STAND" and then switch on the inert gas supply to

the 747 VA Stand by pressing the softkey

. This pressurizes

the multi-mode electrode 38

38 and the mercury begins to flow slowly out of

the capillary.

Gently tap the MME with your finger (to remove any air bubbles) and

allow the mercury to flow out of the capillary into the empty measuring

vessel for approx. 2 min.

Fill measuring vessel 28

28 with 10 mL ultrapure water and add 1 drop KCl

solution (in pure water, mercury drops from the capillary only with diffi-

culty).

Allow mercury to flow out of the capillary for ca. 2 min while checking the

drop formation: The drop time should be ca. 3 s.

4 Adjusting the sealing needle 97

97

Turn slotted screw 96

96 using a suitable coin slowly in a clockwise direction

until the mercury flow stops.

Open slotted screw 96

96 slightly in an anticlockwise direction until the

mercury flow restarts.