279 sdos>m, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 430: Sdos>m

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-279
SDOS>M
Page:
O, S
Stand:
M, A
Abbreviation:
SDOS>M
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. SDOS>M ..... Soln.name ........ Prop.factor ........
Function
Dispense auxiliary solution proportional to the sample size
S0
into the
measuring vessel (MEAS).
Parameter
Range; Default value / Meaning
Soln.name
8 characters; -
Name of the auxiliary solution
The entry of a solution name is mandatory when the auxiliary solution
should be added automatically using a 685 Dosimat, a 700 Dosino or
the 695 Autosampler. In this case the solution name
Soln.name
must
be entered in the parameter of the same name on the "SOLUTIONS"
page and assigned a position
Pos.
(see section 5.3.4).
Without entry of a name for
Soln.name
in the
SDOS>M
instruction,
without entry of this name on the "SOLUTIONS" page or without a
position assignment, the auxiliary solution must be dispensed manually
or with the 6.5611.010 Pipetting equipment (the prompt
DOS
"Soln.name" "V.add" > M
then appears during the program run).
Prop.factor
0.0010...99999.9; 1
Proportionality factor
Specification of the proportionality factor for calculation of the addition
volume
V.add
:
V.add (mL) = Prop.factor * S0 (mL or g)
Description
The
SDOS>M
instruction is used to dispense an auxiliary solution (e.g.
supporting electrolyte, reagent solution) proportional to the sample size
S0
into the measuring vessel (MEAS) on the 747 VA Stand.
Automatic addition can be performed with a 685 Dosimat, a 700 Dosino
or directly with the 695 Autosampler. The appropriate position of the
auxiliary solution must be specified on the "SOLUTIONS" page (see
Soln.name
) for this.
Comments
On execution of the
SDOS>M
instruction with a 685 Dosimat, a 700
Dosino or the 695 Autosampler, a wait is made until the addition is
complete before the instruction following
SDOS>M
is executed.
see also
SDOS>A, SDOS>F, DOS/M, DOS>M, DOS>F, DOS>A, PDOS>M