Metrohm 746 VA Trace Analyzer User Manual
Page 105

4 Introduction to the operation
746 VA Trace Analyzer / 747 VA Stand
4-20
•
First the full report
Report Full
is printed out with all results:
================== METROHM 746 VA TRACE ANALYZER (5.746.0101) ==================
Determ. : 03201614 User: Date: 1997-03-20
Modified : no Run : 1 Time: 16:14:53
Sample table: -
--------------------------------------------------------------------------------
Pos. Ident.1/S1 Ident.2/S2 Ident.3/S3 Method.call Sample size/S0
0.1 mL
--------------------------------------------------------------------------------
Method : TESTPb
Title : Pb determination in ion standard solution
Remark1 :
Remark2 :
--------------------------------------------------------------------------------
Substance : Pb Comments
Mass conc.: 1.032 g/L Mass : 103.2 ug ––––––––––––––––––
MC.dev. : 0.011 g/L (1.05%) Add.mass : 100 ug
Cal.dev. : - V0.sample: 100 uL
VR U/mV I/nA I.mean Std.dev. I.delta Comments
––– ––––– –––––– –––––– ––––––– ––––––– ––––––––––––––––––
00 -355 -195.9 -194.7 1.569
01 -355 -195.3
02 -355 -192.9
10 -356 -385.2 -385.5 0.4942 -190.8
11 -356 -386.1
12 -356 -385.3
20 -357 -568.1 -568.5 1.968 -182.9
21 -357 -566.7
22 -357 -570.6
Substance Techn. Y.reg/offset Slope Nonlin. Mean deviat.
––––––––– –––––––– –––––––––––– –––––––––– –––––––––– ––––––––––––
Pb std.add. -1.957e-07 -3.907e-05 2.137e-09
C# Workg.com.var Remark
–– ––––––––––––– ––––––––––––––––
Final results +/- Res.dev. % Comments
––––––––––––––––––––––––––––––––– ––––––––––––––––– ––––––––––––––––––
Pb = 1.032 g/L 0.011 1.05
•
The full report comprises the following parts:
1
Data of determination
Name, date and time, user, run number, etc.
2
Data of sample
Sample identification, sample size
3
Data of method
Name, title, remarks
4
Data of substance
Substance name, mass concentration found and
evaluation
its absolute and relative total deviation, substance
mass found in measuring vessel, standard addition
mass and sample volume in the measuring vessel
5
Data of peak evaluation
Overview of the 9 individual measurements with
the following results: peak voltage
U, peak current
I, mean value of the peak current I.mean with
associated standard deviation
Std.dev. and
difference between the peak currents
I.delta
6
Data for standard
Current value
Y.reg for the sample calculated
addition
from the standard addition curve, slope and mean
deviation of standard addition curve
7
Common variables
not used in this example
8
Final results
Final result with formula calculation
•
You will find further information regarding the report in section 5.5.2
"RESULTS" and section 5.6.5 "DOCUMENTATION".