5.4.4 SAMPLES

746 VA Trace Analyzer / 747 VA Stand

5-63

Parameter

Range; Default value / Meaning

Method.call

XXXXXXXX

--------

11 characters;

Method call

For automatic call-up of a method, method name, stand number and
program run mode can be entered here. The entry must have the
following format:

xxxxxxxx.SR

xxxxxxxx

=

Method name

S

=

Stand address (

1...3

)

R

=

Run mode (

C,D,T

)

(

Calibration/Determination/

Test

)

With this entry, all methods loaded in the method memory are dis-
played in the softkeys and can thus be taken over directly.

Method name

XXXXXXXX

is copied to the field

Method.call

.

Without entry of a stand number, the VA Stand set in the parameter

Stand address

on the "MONITORING" page is used as default.

Without entry of the run mode, the program run mode set in the
parameter

Run mode

on the "MONITORING" page is used as default.

The automatic method call functions only when the specified method
is present in the working memory.

Sample size/S0

1.0...9999.9 uL...L / ug...kg; - L

Sample size

Entry of the sample size (variable

S0

) in this column as either a volume

or weight is mandatory, otherwise the determination can not be
started (error message

Missing sample

). Entry of an asterisk

*

is also

possible, but only

if no concentrations or dispensed volumes based on

S0

need be

calculated, or

if the instruction

SMPL/F

is used with the 695 Autosampler with

subsequent solution transfer to the auxiliary or measuring vessel,
or

if the instruction

CSOLN>M

or

CSOLN/F

for calibration solutions is

used with the 695 Autosampler.

The use of

S0

depends on whether work is performed with undiluted

or diluted samples:

Undiluted sample, entry as volume
The entered sample size

S0

(in

µ

L, mL or L) is used as addition

volume for the SMPL instructions. The substance content in g/L is
identical to the calculated mass concentration

ρ

s

(

Mass. conc.

,

see section 5.5.2).

Undiluted sample, entry as weight
If the sample size is entered as a weight, the density must be
entered under

S2

in g/mL (with no entry,

S2 = 1.000

is set

automatically). For calculation of the addition volume for the
SMPL instructions, the entered sample size

S0

in g is divided by

the density

S2*1000

in g/L. To obtain the substance content in

g/g, the calculated mass concentration

ρ

s

(

Mass. conc.

, see sec-

tion 5.5.2) is divided by

S2*1000