beautypg.com

Metrohm 746 VA Trace Analyzer User Manual

Page 384

background image

5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-233

Description

The

DSWEEP

instruction (Dummy Sweep) starts the voltage sweep

defined in the parameters. While measured values are recorded during
the sweep, they are not stored.

U

t

U.start

U.end

U.step

t.step

Comments

The electrode type must be defined before the

DSWEEP

instruction.

Admissible are

DME

,

SMDE

,

HMDE

or

RDE

.

The measurement mode must be defined before the

DSWEEP

instruction. Admissible are

DCTMODE

,

DPMODE

,

AC1MODE

,

AC2MODE

or

SQWMODE

, but not

PSAMODE

or

CYCMODE

.

With the

DSWEEP

instruction, the potentiostat always operates in the

operating mode "fast".

If the final voltage

U.end

should not be switched off at the end of the

program, enter a suitable standby voltage

U.standby

in a

ØMEAS

instruction.

see also

SWEEP, FSWEEP, SSWEEP, DME, SMDE, HMDE, RDE, DCTMODE,

DPMODE, AC1MODE, AC2MODE, SQWMODE