Metrohm 746 VA Trace Analyzer User Manual
Page 205

5.4 Storages
746 VA Trace Analyzer / 747 VA Stand
5-54
Parameter
Range; Default value / Meaning
Files
display only
Number of files
Display of the number of stored determinations (maximum 256
determinations can be stored).
Available
display only
Free storage space
Display of the free storage space in bytes.
File
8 characters;
Determination name
In copying a determination from the working memory to the determi-
nation list, the name
MMDDhhmm.dtm
is allocated automatically and
comprises the month (
MM
) – day (
DD
) – hours (
hh
) – minutes (
mm
) when
the determination was recorded and the extension
.dtm
. The first 8-
digit part of the determination name can be changed, however the
extension
.dtm
must be retained as it characterizes determination
files.
Secured determinations (
*
) must first be unsecured with
before their name can be changed.
For the determination name all characters are allowed with the excep-
tion of the character "
.
".
Content
50 characters;
Content information for determination
In copying a determination from the working memory to the determi-
nation list, the three parameters
Sample table
,
S1
and
Method call
are entered automatically if they were available on the "MONITORING"
or "SAMPLES" page at the time of recording.
This 50-character content information can be changed at will and is
retained in direct copying to the data card or the RS interfaces.
However, if the determination is loaded in the working memory and
then copied from there to another destination, this content information
will be lost.
Secured methods (
*
) must first be unsecured with
before
their title can be changed.
Size
display only
Memory size
Display of the memory requirements of the determination in bytes.
Full condition
push out, stop; stop
Mode when memory full
Specification of what should happen when the automatic storage is
switched on (
Det.storage = on
on the "MONITORING" page), but the
determination memory is full:
push out
Create space by deleting the oldest
determination(s)
stop
Discontinue determination series
The mode can be switched with the softkey
.