3.4 Multi-mode electrode (MME)

746 VA Trace Analyzer / 747 VA Stand

3-21

5 Evacuating in vertical position

Place multi-mode electrode 38

38 vertically in the electrode holder 114

114 (see

Fig. 15-1).

Evacuate for ca. 2 min in this position.

6 Evacuating in inclined position

Carefully tilt multi-mode electrode 38

38 in the electrode holder 114

114 to an

inclined position and continue evacuating (see Fig. 15-2).

7 Release vacuum

As soon as mercury issues from the tip of the glass capillary 109

109 into the

filling tubing 115

115, carefully release the vacuum by opening the vacuum

release tap.

The filling tubing 115

115 must not be disconnected from the glass capil-

lary 109

109 when under vacuum, otherwise the mercury which has

issued from the capillary would be sprayed onto the tubing wall and

can no longer be disposed of in drop form!

Tap the glass capillary 109

109 gently by hand so that any mercury drops at

its tip are knocked into the filling tubing 115

115.

Disconnect filling tubing 115

115 with filling cone 116

116 from glass capillary

109

109.

Place multi-mode electrode 38

38 in a horizontal position in the electrode

holder 114

114 (see Fig. 15-3).

From now on, the MME must be left in this position until it is installed in

the stand!

8 Install multi-mode electrode in 747 VA Stand

With measuring head arm 40

40 tilted back, push empty measuring vessel

28

28 into the holder 26

26 of the 747 VA Stand and then lower measuring head

arm 40

40.

Carefully insert multi-mode electrode 38

38 in opening 76

76 of the measuring

head 41

41 (during insertion, the tip of the capillary must not touch the

measuring head) and push in as far as it will go.

9 Connect multi-mode electrode

Screw FEP tubing 51

51 for the inert gas supply into connection 93

93 of the

multi-mode electrode 38

38.

Screw FEP tubing 59

59 for the inert gas supply into connection 94

94 of the

multi-mode electrode 38

38.

Attach electrode cable 37

37 (WE) to screw connection 110

110 of the multi-

mode electrode 38

38: push cable lug under the screw and then tighten

screw firmly.