3.4 Multi-mode electrode (MME)
746 VA Trace Analyzer / 747 VA Stand
3-21
5 Evacuating in vertical position
•
Place multi-mode electrode 38
38 vertically in the electrode holder 114
114 (see
Fig. 15-1).
•
Evacuate for ca. 2 min in this position.
6 Evacuating in inclined position
•
Carefully tilt multi-mode electrode 38
38 in the electrode holder 114
114 to an
inclined position and continue evacuating (see Fig. 15-2).
7 Release vacuum
•
As soon as mercury issues from the tip of the glass capillary 109
109 into the
filling tubing 115
115, carefully release the vacuum by opening the vacuum
release tap.
The filling tubing 115
115 must not be disconnected from the glass capil-
lary 109
109 when under vacuum, otherwise the mercury which has
issued from the capillary would be sprayed onto the tubing wall and
can no longer be disposed of in drop form!
•
Tap the glass capillary 109
109 gently by hand so that any mercury drops at
its tip are knocked into the filling tubing 115
115.
•
Disconnect filling tubing 115
115 with filling cone 116
116 from glass capillary
109
109.
•
Place multi-mode electrode 38
38 in a horizontal position in the electrode
holder 114
114 (see Fig. 15-3).
From now on, the MME must be left in this position until it is installed in
the stand!
8 Install multi-mode electrode in 747 VA Stand
•
With measuring head arm 40
40 tilted back, push empty measuring vessel
28
28 into the holder 26
26 of the 747 VA Stand and then lower measuring head
arm 40
40.
•
Carefully insert multi-mode electrode 38
38 in opening 76
76 of the measuring
head 41
41 (during insertion, the tip of the capillary must not touch the
measuring head) and push in as far as it will go.
9 Connect multi-mode electrode
•
Screw FEP tubing 51
51 for the inert gas supply into connection 93
93 of the
multi-mode electrode 38
38.
•
Screw FEP tubing 59
59 for the inert gas supply into connection 94
94 of the
multi-mode electrode 38
38.
•
Attach electrode cable 37
37 (WE) to screw connection 110
110 of the multi-
mode electrode 38
38: push cable lug under the screw and then tighten
screw firmly.