Metrohm 746 VA Trace Analyzer User Manual

Page 167

background image

5.3 Set Up

746 VA Trace Analyzer / 747 VA Stand

5-16

5.3.1

VA TRACE ANALYZER

VA TRACE

ANALYZER

W

This dialog page is used to set the instrument parameters of the

746 VA Trace Analyzer, the parameters of the RS232 interfaces

and those of the output control lines.

746 VA TRACE ANALYZER

Program 5.746.0101

Date ....:..:.. RS232 settings Ifc.1 Ifc.2

Time ..:..:.. ––––––––––––––––––––––––––––––––––––––––––––

Beeper ... Baud rate .... ....

Data bits .. ..

Display Stop bit(s) .. ..

––––––––––––––––––––––––––––– Parity ..... .....

Switch–off time ... min Handshake ........ ........

Contrast XXX Receive mode ... ...

Configuration Address Balance type ............

––––––––––––––––––––––––––––– Printer type ............ ............

XXXXXXXXXXXXXXXXXXX X Printer width ........ ........

XXXXXXXXXXXXXXXXXXX X Remote keyboard ........

XXXXXXXXXXXXXXXXXXX X

XXXXXXXXXXXXXXXXXXX X Ctrl.lines # 87654321 Printer

XXXXXXXXXXXXXXXXXXX X ––––––––––––––––––––––– ––––––––––––––––––

Scan XXXXXXXX Type ............

Set ........

Ext. start ...

Modify

column

Set Ctrl

lines

Contrast

up

Contrast

down

Reset

Ifc.1

Reset

Ifc.2

Save

Ifc.par.

Diagn.

Softkeys

Meaning

Modify

column

Modify parameters column-by-column

Set Ctrl

lines

Commands for 708 Sampling Unit

For manual operation via control lines of the 708 Sampling Unit
connected to the 746 VA Trace Analyzer with a 6.2143.100 or
6.2143.110 cable, the following commands are available, which
change the last three digits of the parameter

Set

simultaneously:

FILL

PUMP ON

Set injection valve to "FILL", switch pump on

(

Set = *****101

)

FILL

PUMP OFF

Set injection valve to "FILL", switch pump off

(

Set = *****100

)

INJECT

PUMP ON

Set injection valve to "INJECT", switch pump on

(

Set = *****011

)

INJECT

PUMP OFF

Set injection valve to "INJECT", switch pump off

(

Set = *****010

)

Contrast

up

Increase screen contrast

The screen is set one level brighter (+2). The current contrast value is
displayed in the

Contrast

parameter.

Continuous brightening can be achieved by pressing the keys
and <é> simultaneously.