Metrohm 746 VA Trace Analyzer User Manual
Page 232

5.5.1 MONITORING
746 VA Trace Analyzer / 747 VA Stand
5-81
Parameter
Range; Default value / Meaning
XXXXXXXX
--------
Method name
XXXXXXXX
is copied to the field
Method.call
.
Without entry of a stand number, the VA Stand set in the parameter
Stand address
on the "MONITORING" page is used as default.
Without entry of the run mode, the program run mode set in the
parameter
Run mode
on the "MONITORING" page is used as default.
The automatic method call functions only when the specified method
is present in the working memory.
Sample size/S0
1.0...9999.9 uL...L / ug...kg; - L
Sample size
Entry of the sample size (variable
S0
) in this column as either a volume
or weight is mandatory, otherwise the determination can not be
started (error message
Missing sample
). Entry of an asterisk
*
is also
possible, but only
•
if no concentrations or dispensed volumes based on
S0
need be
calculated, or
•
if the instruction
SMPL/F
is used with the 695 Autosampler with
subsequent solution transfer to the auxiliary or measuring vessel,
or
•
if the instruction
CSOLN>M
or
CSOLN/F
for calibration solutions is
used with the 695 Autosampler.
You will find a detailed description of the
Sample size/S0
parameter
in section 5.4.4 "SAMPLES".