5.5.1 MONITORING

746 VA Trace Analyzer / 747 VA Stand

5-81

Parameter

Range; Default value / Meaning

XXXXXXXX

--------

Method name

XXXXXXXX

is copied to the field

Method.call

.

Without entry of a stand number, the VA Stand set in the parameter

Stand address

on the "MONITORING" page is used as default.

Without entry of the run mode, the program run mode set in the
parameter

Run mode

on the "MONITORING" page is used as default.

The automatic method call functions only when the specified method
is present in the working memory.

Sample size/S0

1.0...9999.9 uL...L / ug...kg; - L

Sample size

Entry of the sample size (variable

S0

) in this column as either a volume

or weight is mandatory, otherwise the determination can not be
started (error message

Missing sample

). Entry of an asterisk

*

is also

possible, but only

if no concentrations or dispensed volumes based on

S0

need be

calculated, or

if the instruction

SMPL/F

is used with the 695 Autosampler with

subsequent solution transfer to the auxiliary or measuring vessel,
or

if the instruction

CSOLN>M

or

CSOLN/F

for calibration solutions is

used with the 695 Autosampler.

You will find a detailed description of the

Sample size/S0

parameter

in section 5.4.4 "SAMPLES".