5-217
CSOLN>M
Page:
O, S
Stand:
M, A
Abbreviation:
CSOLN>
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. CSOLN>M ..... addition into MEAS
Function
Dispense calibration solution into the measuring vessel (MEAS).
Description
The
CSOLN>M
instruction is used for the recording of calibration curves
with the techniques
cal.crv.
and
int.std
. Within a CAL loop, it defines
that the calibration solution should be dispensed into the measuring
vessel.
The calibration solutions can be dispensed manually, with the
6.5611.010 Pipetting equipment, or via the 695 Autosampler, but not with
a 685 Dosimat or 700 Dosino. On manual addition with the
CSOLN>M
instruction the prompt
CSOLN "CSoln.name" "Mass conc." > M
appears
showing the name and concentration of each calibration solution. In
additions with the 695 Autosampler, the calibration solutions are
transferred in turn to the measuring vessel from the sample rack.
Comments
•
The number of calibration solutions is defined with
n
in the
CAL)n
instruction. A valid mass concentration
Mass conc.
must be entered
on the "SUBSTANCES/Calibration" page for each of these solutions.
On manual addition, a name should also be entered for each solution
in the parameter
CSoln.name
which then appears in the message for
addition.
•
The addition volume valid for all calibration solutions must be entered
in the
V0.CSoln
parameter on the "SUBSTANCES/ Calibration" page.
•
If work is performed with
Auto.samples = on
, on the "SAMPLES"
page sample data lines must be entered in immediate succession for
all n calibration solutions with the following holding for their parame-
ters:
Pos
.
Position number on the sample rack
(freely selectable, only with 695 Autosampler)
S1...S3
Sample identifications (only entries of the last
calibration solution are used and stored)
Method.call
Name of the method +
.C
(for mode
calibration
)
Sample Size/S0
Entry of an asterisk
*
(value not used, but field must not be blank)
•
If the entire analysis solution is changed each time following the
recording of the calibration curve, a
RINSE
instruction must be in-
serted in the CAL loop which resets the volume bookkeeping in the
measuring vessel to 0 (see also
RINSE). If its parameter
Cycles > 0
is set, the measuring vessel is also rinsed by means of the two
rinsing Dosinos connected to the 746 VA Trace Analyzer or the 695
Autosampler. If no rinsing Dosinos are connected, the prompt
Change measuring vessel
to change the solution appears.
•
Without a
RINSE
instruction, the volume in the measuring vessel is
added continuously. This can be utilized for the recording of calibra-
tion curves in which the calibration solutions are prepared each time
by manual addition of a concentrated standard solution to the analy-
sis solution.