5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-217

CSOLN>M

Page:

O, S

Stand:

M, A

Abbreviation:

CSOLN>

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. CSOLN>M ..... addition into MEAS

Function

Dispense calibration solution into the measuring vessel (MEAS).

Description

The

CSOLN>M

instruction is used for the recording of calibration curves

with the techniques

cal.crv.

and

int.std

. Within a CAL loop, it defines

that the calibration solution should be dispensed into the measuring
vessel.

The calibration solutions can be dispensed manually, with the
6.5611.010 Pipetting equipment, or via the 695 Autosampler, but not with
a 685 Dosimat or 700 Dosino. On manual addition with the

CSOLN>M

instruction the prompt

CSOLN "CSoln.name" "Mass conc." > M

appears

showing the name and concentration of each calibration solution. In
additions with the 695 Autosampler, the calibration solutions are
transferred in turn to the measuring vessel from the sample rack.

Comments

The number of calibration solutions is defined with

n

in the

CAL)n

instruction. A valid mass concentration

Mass conc.

must be entered

on the "SUBSTANCES/Calibration" page for each of these solutions.
On manual addition, a name should also be entered for each solution
in the parameter

CSoln.name

which then appears in the message for

addition.

The addition volume valid for all calibration solutions must be entered
in the

V0.CSoln

parameter on the "SUBSTANCES/ Calibration" page.

If work is performed with

Auto.samples = on

, on the "SAMPLES"

page sample data lines must be entered in immediate succession for
all n calibration solutions with the following holding for their parame-
ters:

Pos

.

Position number on the sample rack
(freely selectable, only with 695 Autosampler)

S1...S3

Sample identifications (only entries of the last
calibration solution are used and stored)

Method.call

Name of the method +

.C

(for mode

calibration

)

Sample Size/S0

Entry of an asterisk

*

(value not used, but field must not be blank)

If the entire analysis solution is changed each time following the
recording of the calibration curve, a

RINSE

instruction must be in-

serted in the CAL loop which resets the volume bookkeeping in the
measuring vessel to 0 (see also

RINSE). If its parameter

Cycles > 0

is set, the measuring vessel is also rinsed by means of the two
rinsing Dosinos connected to the 746 VA Trace Analyzer or the 695
Autosampler. If no rinsing Dosinos are connected, the prompt

Change measuring vessel

to change the solution appears.

Without a

RINSE

instruction, the volume in the measuring vessel is

added continuously. This can be utilized for the recording of calibra-
tion curves in which the calibration solutions are prepared each time
by manual addition of a concentrated standard solution to the analy-
sis solution.