5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-293

Parameter

Range; Default value / Meaning

Meas.cycles

1...254; 4

Number of measurement cycles

Number of square wave voltage cycles with current measurement
which are performed for every voltage step.

The following condition holds for the definition of

Meas.cycles

:

32 ms

Meas.cycles < ––––––––

t.meas

Description

With the

SQWMODE

measurement mode, a square wave alternating

voltage of small, constant amplitude and low frequency is superimposed
on a stepwise rising direct voltage ramp. Measurement of the current I as
a function of the voltage U involves several determinations of the
differences between the currents at each maximum and minimum
square wave voltage. The nature of the measurement is illustrated below
using the SMDE as an example.

Adjustable parameters:

SQWMODE

t.step
t.meas
U.ampl
Modul.freq.
Prep.cycles
Meas.cycles

SWEEP

U.start
U.end
U.step

t.drop

t

U

t

U.step

U.start

U.ampl

t.step

t.modul = 1/Modul.freq.

F (Hg)

Prep.

cycles

Meas.

cycles

t.meas (I

n1

)

t.meas (I

n2

)

Area (Hg)